for dedicated SPECT/PET for brain, newborn babies and small animal imaging with multiplied patient throughput
More heads - better and faster imaging The Coincidence (PET) detection offered as an option. More patients within less time. The image acquisition time is half of that of a dual head SPECT system. High resolution imaging. Using low energy ultra high resolution collimators the special detectors result in excellent system resolution in the low gamma energy region. |
| Detectors |
Four rectangular high stability detectors assembled with high optical and mechanical quality
Detector electronics Compact, highly integrated, one-board easily serviceable construction without tuning potentiometers:
|
| Imaging Table |
Integrated imaging table for SPECT and whole body examinations
Gantry
Gantry monitor continuously presents information about gantry position, system function and indicates service instructions. Optional accessories
|
| Technical Specifications |
Measured per NU-1, 1994 using Tc-99m where applicable. |
Collimators
|
| Performance characteristic | Nucline™ X-Ring | Units | |||||
|---|---|---|---|---|---|---|---|
| SPECT UFOV (Useful Field of View) | 215 x 230 |
mm |
|||||
| Number of PMTs | 4 x 60 |
4 x 33 |
|
||||
| Crystal thickness | 6.4 |
9.5 |
12.5 |
mm |
|||
| Intrinsic spatial resolution | CFOV | FWHM | 2.3 |
2.9 |
3.4 |
3.9 |
mm |
| FWTM | 4.6 |
5.8 |
6.8 |
7.8 |
mm |
||
| UFOV | FWHM | 2.4 |
3.0 |
3.5 |
4.0 |
mm |
|
| FWTM | 4.8 |
6.0 |
7.0 |
8.0 |
mm |
||
| Intrinsic energy resolution | UFOV | FWHM | 9.7 |
|
|||
| Intrinsic flood field uniformity | CFOV | Differential | 1.9 |
% |
|||
| integral | 2.4 |
% |
|||||
| UFOV | differential | 2.4 |
% |
||||
| integral | 2.9 |
% |
|||||
| Intrinsic spatial linearity | CFOV | differential | 0.18 |
mm |
|||
| absolute | 0.38 |
mm |
|||||
| UFOV | differential | 0.20 |
mm |
||||
| absolute | 0.40 |
mm |
|||||
| Intrinsic count rate performance without scatter | maximum | 200 |
kcps |
||||
| at 20% loss | 160 |
kcps |
|||||
| System spatial resolution with LEGP / LEHR collimator | FWHM | 8.3 / 6.6 |
8.5 / 6.9 |
8.7 / 7.1 |
8.9 / 7.2 |
mm |
|
| FWTM | 16.1/13.1 |
16.5/13.5 |
16.7/13.7 |
16.9/13.9 |
mm |
||
| Maximal matrix size | 1024 x 1024 |
pixel |
|||||
| Camera weight | 800 |
kg |
|||||
| Minimal room size | 12 |
m2 |
|||||
| Uptime warranty | 98 |
% |
|||||
| Movements directions | 2 |
|
|||||
| Collimator mounting | slide-in, bolt fixing |
|
|||||
| Collimator changing time | < 1.5 |
min |
|||||
| Acquisition Console and Gamma Camera Electronics |
Ergonomic
acquisition workstation console stands on wheels. Full-digital electronics assembled from the latest "high-tech" elements including fast PCI bus acquisition interface.
|
| Typical Room Layout |
| The minimum size (4.2 x 2.8 m – 13.78 x 9.2 ft) room layout, and space requirement for brain SPECT examination (not drawn to scale). Layout Only - Not For Construction. Click on the image to enlarge it. |
|