MT-ERI-A10 EXRADIN A10 ELECTRON BEAM CHAMBERFeatures:
  • Similar to Markus
  • Fully guarded design
  • Planar geometry
  • Small collecting volume
  • Homogeneous construction
  • Thin stretched conductive window
  • Waterproofing cover for TG-51 or TG-39 available
  • Low mass collector
  • Integral flexible low-noise cable
  • Handsome wooden case
Collecting Volume: 0.051 cm3
Nominal Calibration Factor: 60 R/nC
Centroid of Collecting Volume: 1.0 mm from inside window surface
Collector Diameter: 5.4 mm
Window-Collector Gap: 2.0 mm
Window: Conductive Kapton film, 3.86 mg/cm2
Window Support Rings, Collector, and Guard Material: Model A10 – Shonka air-equivalent plastic C552
Maximum Polarizing Potential: Greater than 1000 V
Inherent Leakage Currents: Less than 10-15 Amperes
Cable: 50 Ohms, 29pF/f, 3 m long
Signal Connector: Choice of coaxial BNC and triaxial BNC and TNC plugs and jacks.
High Voltage Connector: Integral with triaxial connectors (window of chamber is common with connector body). Banana, MHV, and SHV plugs are also available.
Venting: Chamber is vented at connector via the triaxial cable.
Stem: Integral with chamber. Black delrin, 0.350" OD by 3" long. Other lengths are available upon request.