Collecting Volume: |
0.62 cm3 |
|
Nominal Calibration Factor: |
5.5 R/nC |
|
Centroid of Collecting Volume: |
2.0 mm from surface of window |
|
Collector Diameter: |
20.0 m |
|
Window Thickness: |
1.0 mm |
|
Window-Collector Gap: |
2.0 mm |
|
Body, Collector, and Guard Material: |
A11: Shonka air-equivalent plastic C552
P11: Polystyrene equivalent plastic D400
T11: Shonka tissue-equivalent plastic A150 |
|
Maximum Polarizing Potential: |
Greater than 1000 V |
|
Inherent Leakage Currents: |
Less than 10-15 Amperes |
|
Cable: |
50 Ohms, 29pF/f, 3 m long |
|
Signal Connector: |
Choice of coaxial BNC and triaxial BNC and TNC plugs and jacks. |
|
High Voltage Connector: |
Integral with triaxial connectors (shell of chamber is common with connector body). Banana, MHV, and SHV plugs are also available. |
|
Vent Tubes: |
Two 21 Ga stainless steel tubes which may be extended to connector via 22 Ga TFE tubing for possible immersion in liquid phantoms. |
|
Stem: |