Unfors DENT X-ray System 10 is made for kVp, dose, rate, time and pulse measurements in dental X-ray.
If you are a dental service engineer, Unfors Mult-O-Meter 512L is a perfect instrument for measuring all types of dental X-ray equipment. This model is used by many
major dental X-ray manufacturers for field work as well as factory calibrations. The Unfors Mult-O-Meter has a narrow detector which can be used on both panoramic and
intra oral. The unique magnetic OPG holder positions the kVp and dose sensors in a straight line to align themselves up to the configuration of the beam. The background
fluorescent screen makes it easy to locate the actual beam. With the Unfors DAQ-10 software you easily make reports directly in Excel. Unfors Mult-O-Meter 512L with external detector for kVp DENT, Dose DENT, Rate DENT, Time, Pulse and Extended Functions The Unfors X-ray Systems offer a cost effective and rapid way for
measurements. The Systems are designed with the aim to make the measurement procedure simple and straightforward. All Unfors X-ray Systems comes in a customized case
with a Quick Guide and a comprehensive CD-manual in. The Systems includes Extended Functions and an IR (infrared) link for transmission of measured data to a PC and the
Unfors DAQ-10 software for Excel. |
| Specifications: |
| kVp DENT | Range (auto): 45 - 110 kVp Minimum exposure: 3 mA at 70 kVp, 50 cm |
| Dose DENT | Range (auto): 8 µGy- 9999 Gy, (1 mR - 9999 R) Energy dependence: < 5 % at 2.5 mm Al, (50 - 150 kVp) |
| Rate DENT | Range (auto): 8 µGy/s - 80 mGy/s, (55 mR/min - 550 R/min) Energy dependence: < 5 % at 2.5 mm Al, (50 - 150 kVp) |
| Time | Range (auto): 1 ms - 9999 s measured with the trigger detector Inaccuracy: 0.5 % or 3 digits |
| Pulse | Range: 1 ms - 9999 s measured with the trigger detector Inaccuracy: 1 pulse |
| Extended functions | Normalization of dose Dose reproducibility mAs linearity Simplified HVL Time corresponding to the 75 % triglevel of kVp waveform kVp with a 5, 50, 150 ms delay |
| IR Link | Equipped with an infrared interface for transmission of measured data to a PC. The IR receiver connected to the serial port of a PC, receives the measured data. For data collection and calculations, the Unfors Excel Add-in software is available for user customized reports. |
| Unfors DAQ-10 Software | Operating system Windows 98/ME, Windows 2000, Windows XP Microsoft Excel Office 97, 2000 and XP |
| OPG Holder magnetic | Magnetic holder with fluoroscent screen for easy positioning of Unfors Mult-O-Meter external detector on panoramic dental. |
| Case | Unfors QA Case 2 |
| Delivery | 7-10 days, ARO |