| Manufacturers | Radiation Oncology | Medical Physics | Diagnostic Imaging | Radiation Safety | Exercise Systems |
Unfors R/F X-ray System 61 is a perfect system when you need a small and truly portable solution for
field service work when a wide dose range is not needed. The Unfors Mult-O-Meter 706L measures kVp, dose, rate, dose/frame, time, mA and mAs simultaneously on most R/F
machines. Together with the System components it provides a user-friendly all-in-one solution for service work - in a size smaller than a regular mAs meter.Unfors Mult-O-Meter 706L with external detector for kVp R/F, Dose R/F, Rate R/F, Time and invasive detector for mA, mAs and Time and Dose/Frame, IR Link and Extended Functions The Unfors X-ray Systems offer a cost effective and rapid way for measurements. The Systems are designed with the aim to make the measurement procedure simple and
straightforward. All Unfors X-ray Systems comes in a customized case with a Quick Guide and a comprehensive CD-manual in. The Systems includes Extended Functions and an
IR (infrared) link for transmission of measured data to a PC and the Unfors DAQ-10 software for Excel. |
| Specifications: |
| kVp R/F | Range (auto): 45 - 155 kVp Minimum exposure: 3 mA at 70 kVp, 50 cm |
| Dose R/F | Range (auto): 8 µGy- 9999 Gy, (1 mR - 9999 R) Energy dependence: < 5 % at 2.5 mm Al, (50 - 150 kVp) |
| Rate R/F | Range (auto): 8 µGy/s - 80 mGy/s, (55 mR/min - 550 R/min) Energy dependence: < 5 % at 2.5 mm Al, (50 - 150 kVp) |
| Time | Range (auto): 1 ms - 9999 s measured with the trigger detector Inaccuracy: 0.5 % or 3 digits |
| Dose/Frame | Range (auto): 200 nGy - 9999 Gy, (23 mR - 9999 R) Minimum pulses/s: 10 Minimum pulse width: 1 ms |
| Invasive mA/mAs | Range mA (auto): 0.3 - 2000 mA Inaccuracy: 2 % or 0.05 mA Range mAs (auto): 0.10 - 9999 mAs Inaccuracy: 2 % |
| Extended functions | Normalization of dose Dose reproducibility mAs linearity Simplified HVL Time corresponding to the 75 % triglevel of kVp waveform kVp with a 5, 50, 150 ms delay |
| IR Link | Equipped with an infrared interface for transmission of measured data to a PC. The IR receiver connected to the serial port of a PC, receives the measured data. For data collection and calculations, the Unfors Excel Add-in software is available for user customized reports. |
| Unfors DAQ-10 Software | Operating system Windows 98/ME, Windows 2000, Windows XP Microsoft Excel Office 97, 2000 and XP |
| Pen detector holder | Extendable holder for Unfors Mult-O-Meter external detectors for easy positioning on I.I.'s, in buckys etc. |
| Case | Unfors QA Case 2 |
| Delivery | 7-10 days, ARO |
| Harpell Associates Inc., |
| 1272 Speers Road, Unit 2, Oakville, Ontario, Canada, L6L 2X4 |
| Phone (905) 825-2588, Fax (905) 825-0234, Toll Free: 1-800-387-7168 |
| Email: Info@Harpell.ca |