Extremely Intelligent QA DeviceThe Unfors Xi is a new generation, multi-function, intelligent X-ray meter. Unfors is renowned for its pocket-sized and easy-to-use meters that improve productivity. Now the Unfors Xi adds even more user friendliness with Active Compensation for all beam qualities. |
You have five minutes to find the problem. What are your expectations of your QA device? File format: Flash |
No correction is needed for varying total filtration. Watch the video. File format: Flash |
| Unifors Xi Features: |
One meter - all modalitiesActive Compensation for all beam qualities No set-up time Easy to use |
| Active Compensation Active Compensation is a new proprietary feature of the Unfors Xi. Utilizing multiple sensors and advanced calculations, the beam quality is automatically determined thereby eliminating the need for further corrections of measured kVp and dose values. This built-in intelligence enhances ease-of-use. |
![]() |
| “All-In-One” “All-In-One” is a perfect descriptor for the Unfors Xi since one detector can simultaneously measure kVp, dose, dose rate, pulses, pulse rate, dose/pulse, time and waveforms for all modalities. Three versions of Xi Detectors are available: R/F + Mammo, R/F only and Mammo only. |
![]() |
| Pulsed fluoroscopy Pulsed fluoroscopy is used more in today’s radiology departments than ever before. The Unfors Xi makes it very easy to measure any type
of pulsed fluoroscopy since automatic features detect and calculate the pulse rate and dose/pulse. No user interaction is needed. |
![]() |
| Improve productivity “The need for a compact and easy-to-use design was an obvious conclusion drawn from our many conversations with field service
engineers, medical physicists and government inspectors. Our engineers have focused on minimizing the size of the Unfors Xi while maximizing its performance. We provide
you with a small and lightweight, yet powerful QA device which requires minimal training and is easy to use. We help you improve productivity.” |
![]() |
| A complete solution The Unfors Xi is a complete QA system and can be fully upgraded as new features are introduced. |
|
| The Unfors Xi Base Unit The three row alphanumerical display clearly presents all measured parameters. All functions are accessed by pressing one of the
Unfors Xi’s two buttons. The Unfors Xi operates on a rechargeable 9V battery. |
![]() |
![]() |
The Unfors Xi Detector Dose measurements with the Unfors Xi are extremely accurate thanks to the Active Compensation feature. Traditional bulky and
fragile ion chambers can now be substituted with small solid state detectors to achieve accurate measurements. Standards in dose measurements are evolving. Evolve
instantly with a meter that uses tomorrow’s technology. |
| Unfors Xi Configurations: |
Three versions of interchangable Unfors Xi Detectors are available: R/F + Mammo, R/F only and Mammo only. The
Unfors Xi Mammo detector handles the seven most common beam qualities used in today's mammographic X-ray machines. |
Unfors Xi Base Unit includes:
|
Unfors Xi Options:
|
| Unfors Xi R/F & MAM Specifications |
| GENERAL | |
| EMC tested | According to EN 61-000-4-2, EN 61-000-4-3 and EN 55011 |
| Waveform | Single phase to high frequency and pulses with less than 2 s dead time interval. |
| Exposure needed | One |
| Power off | Automatic after 5, 20 or 60 min of inactivity |
| Reset | Automatic |
| Power source | Rechargeable NiMH standard 9V |
| Battery time | 20 hours (11 h with Bluetooth) |
| Read out | Three row alphanumerical |
| Size Base unit | (HxWxL) 28 x 74 x 142 mm, (1.1 x 2.9 x 5.6 in) |
| Detector | (HxWxL) 12 x 22 x 117 mm, (0.5 x 0.9 x 4.6 in) |
| Weight Base unit | 250 g, (8.75 oz) |
| Weight Detector | 50 g, (1.75 oz) |
| kVp | |
| Range R/F (auto) | 35 - 160 kVp/kV |
| Range MAM (auto) | 22 - 40 kVp/kV |
| Accuracy | ± 2 % |
| Sensitivity R/F low | 0.04 mA at 40 kV, 50 cm |
| Sensitivity R/F high | 0.8 mA at 70 kV, 50 cm |
| Sensitivity MAM | 10 mA at 28 kV, 65 cm |
| Dose | |
| Range (auto) | 10 nGy - 9999 Gy, (1 µR - 9999 R) |
| Accuracy R/F | ± 5 % (40 - 150 kVp and for HVL = 1.5 - 14 mm Al*, Active Compensation)or ± 10 nGy (1 µR) |
| Accuracy MAM | ±5 % (22 - 40 kVp and for HVL up to 2.5mm Al added filtration,Active Compensation) |
| Dose rate | |
| Range R/F low (auto) | 10 nGy/s - 1 mGy/s (70 µR/min - 7 R/min) |
| Range R/F high (auto) | 20 µGy/s -1000 mGy/s (140 mR/min - 7000 R/min)at 70 kVp |
| Range MAM (auto) | 10 µGy/s - 100 mGy/s (70 mR/min - 700 R/min) |
| Exposure time | |
| Range (auto) | 1.000 ms - 999.9 s |
| Accuracy | ± 0.5 % or 0.2 ms |
| Pulse | (R/F only) |
| Range (auto) | 1 - 9999 ± 1 pulses |
| Frame rate | (R/F only) |
| Range (auto) | 0.5 - 120 pulses/s |
| Dose per frame | (R/F only) |
| Range (auto) | 1 nGy (0.1µR)-9999 Gy(R)/pulse |
| HVL | |
| Range R/F | 1.00 - 14.0 mm Al |
| Range MAM | 0.2 - 1.2 mm Al depending on beam quality |
| Accuracy R/F | ± 10% or ± 0.2 mm Al (within 1 ‰ of max dose rate for selected sensor) |
| Accuracy MAM | ± 5% (for HVL up to 2.5 mm Al added filtration to each beam quality) |
| Beam qualities | MAM Mo/Mo, Mo/Rh, Mo/Al, Rh/Rh, Rh/Al, W/Al and W/Rh |